Ultrawave PRESSURE LEAK TEST
Load the vessel with 150ml H2O No samples rack.
Set the following testing program:
- Chiller switched on
- cooling water at 10°C (±5°C)
- vessel cooling activated at 25°C
- Pressure release activated, below 80°C at 10bar/min
- Initial pressure load: 40bar
After test save the run on the terminal (LeakTest+ date)