Ultrawave MICROWAVE POWER TEST
Fill the main vessel with Basic Load 150ml H2O + 2ml H3PO4 (or H2SO4). No samples rack.
Set the following testing program:
- Chiller switched on
- cooling water at 10°C (±5°C)
- vessel cooling activated at 25°C
- Pressure release activated, below 80°C at 10bar/min
- Initial pressure load: 30bar
After test save the run on the terminal (Powertest + date)