Telnr. +31(0)346 269090
SK-15 rotor General working specifications
Sample amount
Maximum sample amount: 1 g (of organic dried matter)
VOLUMES
Minimum volume
•10 mL (without inserts)
•5 mL (with QS-50 quartz inserts)
•1 mL (with 3-positions rack inserts)
Maximum working volume
•50 ml (without inserts)
•15 mL (with QS-50 inserts)
•3 mL (with 3-positions rack inserts)
Temperature
Maximum working temperature: 230°C
Maximum holding time at maximum working temperature: 30 minutes
Pressure
Maximum working pressure: 100 bar
Acid solutions
The possibility of use different types of acid solutions is strictly related to the material of the vessels and accessories.
TFM vessels and TFM inserts: HNO3 , HCl, HF, HBF4, H2SO4, H3PO4, HClO4, H2O2
Glass and quartz inserts: HNO3 , HCl, HBF4, H2SO4, H3PO4, HClO4, H2O2